Products
Categories
- Home
- Specialty Gas Equipment
- Regulators
- Modified Regulators "-S"
- 3570-S Modified Single-Stage High-Purity Miniature Regulator - SS
Product Description
Compact, high-purity regulators for use with small cylinders.
Applications
• Applications requiring high-purity gases and a compact regulator due to space limitations
MATHESON strongly recommends the use of a cross-purge with any regulator put into highly corrosive or toxic gas service. You can easily add a cross-purge to this model regulator by selecting "Yes" for the Cross Purge option above.
Design Features
• High-purity 316L stainless steel barstock body
• High-purity 316 stainless steel diaphragm
• 1-1/2” inlet and delivery pressure gauges
• Porous metal filter protects seat from contaminants
• Panel mountable
• 0-3000 psi cylinder pressure gauge
Part Number* | Max Delivery Pressure |
Delivery Pressure Gauge |
Cylinder Pressure Gauge |
SEQ3571CGA-S | 30 psi | 0-60 psi | 0-3000 psi |
SEQ3572CGA-S | 60 psi | 0-100 psi | 0-3000 psi |
SEQ3573CGA-S | 100 psi | 0-200 psi | 0-3000 psi |
*Note: where CGA is replaced by the actual CGA number. Some CGA limitations may apply. (NI) for 1/4" FNPT inlet
Materials of Construction
• Gauges: 316 stainless steel
• Body: 316L stainless steel
• Diaphragm: 316 stainless steel
• Seat: Kel-F 81
• Seals: Teflon
Specifications
• Maximum Inlet Pressure: 3000 psi (20,700 kPa)
• Maximum Flow Rate: 11 SCFH (5 SLPM)
• Flow Capacity (Cv): 0.06
• Operating Temperature: -40°F to 140°F (-40°C to 60°C)
• Porting (Regulator Body): 1/4” NPT Female
• Gauge Ports: 1/8” NPT Female
• Porting Configuration: 2 Low, 2 High
Options
Valve
The standard 3570 Series does not come with an outlet valve. The diaphragm valve is the 4370 Series.
No Lube
A small amount of polychlorotrifluoroethylene is added to the threads to facilitate assembly. For ECD, MS or AED applications where method detection limits are sub-ppm, it is recommended to use a No Lube (-NL) model.
Cross Purge
• Eliminates the escape of hazardous gases when changing cylinders
• Reduces danger of contamination from atmospheric moisture and oxygen
• Minimizes the amount of gas lost during cylinder changes
• All stainless steel construction
• 4360 Series Quarter-Turn Diaphragm Valves used throughout